Trouble Shooting
Should problems arise with the use of the
the recommended solution. If the fault is not found and/or the solution does not correct the problem contact
your local authorized dealer for service immediately.
Fault
The airline tubing that connects the
hand control to the lift has become
disengaged.
The hand control buttons do not op-
erate according to their designations
(e.g. the UP button initiates a down
movement).
The lift does not operate up or down
even when the airline has been
properly connected.
The GREEN light on the left front
side of the lift is ON and the lift does
not operate in the DOWN direction.
The red indicator light on the left
front side of the lift turns RED and/or
a loud alarm sound is heard when an
individual is raised.
One side of the lift tape (strap) is
starting to fray after continued use.
The lift does not pass through a track
component such as a turntable or
gate.
P600 Lift- User Guide (753102)
P-600
lift review the following chart. Find the fault and complete
Recommended Solution
Refer to the section of this manual titled "Attaching the airline tube to the lift".
If this does not correct the problem then contact your local authorized dealer immediately so
that the lift can be checked to ensure proper continued operation.
The airline tubing has not been connected correctly. Refer to the section of this manual titled
"Attaching the airline and hand control to the lift". If this does not correct
the problem then contact your local authorized dealer immediately so that the lift can be
checked to ensure proper continued operation.
The indicator light on the front right side of the lift should be GREEN. Press the ON/OFF
button or UP/DOWN arrow buttons on the hand control. This should activate the lift and the
indicator light turn GREEN.
If the lift still does not function, then the batteries may be low and require charging. Refer to
the section of this manual titled "Charging the lift". Charge the lift for at least one hour
and then try to raise and lower it.
There is a built-in slack tape detector in the lift. This may be sensitive. Take hold of the lift
strap about 10 centimetres above the top of the lift and pull it upwards in order to tighten the
lift strap, and then press the DOWN button. If this corrects the problem temporarily but not
permanently then contact your local authorized dealer so that the lift can be checked to en-
sure proper continued operation.
The batteries are low and require charging. Refer to the section of this manual titled
"Charging the lift". Charge the lift for at least one hour and then try to raise and lower
it.
If this does not correct the problem then contact your local authorized dealer immediately so
that the lift can be checked to ensure proper continued operation.
Check to be sure that the lift is always directly above the individual being lifted. Refer to the
section titled "Basics in transferring an individual" for correct lift positioning. If
fraying still continues then contact your local authorized dealer immediately so that the lift
can be checked to ensure proper continued operation.
Refer to the "Owners Manual" for the specific piece of equipment in question. If the recom-
mended solution does not correct the problem then contact your local authorized dealer im-
mediately so that the lift can be checked to ensure proper continued operation.