Component List; Specifications - Handicare 960-117 Owner's Manual

Table of Contents

Advertisement

Component List

The following components are included with your new P-600 lift system:
P-600 lift

Pneumatic Hand Control
Lift Charger
Owner's Manual
Warranty Card
SLINGS: If a sling has been supplied with the lift refer to the instructions included with the sling.
ACCESSORIES: If additional accessories such as a turntable, or gate system, have been supplied with the lift refer to the
instructions included with those items.
IMPORTANT: Before initial use, the lift unit must be charged for 5 hours. Refer to section titled "Charging Instructions".
The hand control airline tube must also be connected to the lift. If it is not connected refer to the section titled "Attaching the
airline to the lift".
Specifications of P-600 lift
Lift Motor:
Charger Alternate 1 Model:
Charger Alternate 1 Input:
Charger Alternate 1 Output:
Charger Alternate 2 Model:
Charger Alternate 2 Input:
Charger Alternate 2 Output:
Batteries:
Lift Case:
Hand Control:
Lifting Range:
Lift Weight:
Maximum Load:
Internal Fuses:
Duty Cycle:
Rated Performance:
Service Life:
P600 Lift- User Guide (753102)
24 VDC
Soneil 2403SRM30
100-240 VAC, 1.5 Amps, 50-60 Hz
24 VDC, 1.5 Amps
Soneil 2403SRM20
100-240 VAC, 1-0.6 Amps, 50-60 Hz
24 VDC, 1.0 Amp
24 VDC (2 x 12 VDC) 5.0 AH
Flame Retardant ABS
Pneumatic – Piston Displacement
Up to 1.96m
23 lbs (10.432 kg)
600 lbs (274.28 kg)
15 Amp
1 min "ON", 9 mins "OFF"
5-10 lifts at 600 lbs, 10% duty cycle, each lift being 24 inches at the
middle of the lifting range (from 54" strap out to 30" strap out) per full battery
7 years or 22,500 cycles, whichever comes earlier
As a precautionary measure, the lifting strap should be replaced every 5000 cycles or 3 years,
whichever comes earlier. Higher usage lifts may necessitate more frequent replacement of the
lifting strap; please refer to page 20 for General Inspection and Maintenance information.
Rev: 14 MAR 2017
Models Table for P-600 Lifts
Code
Description
303090 P-600 lift including charger and pneumatic
hand control
Shipping/Storage Conditions:
Temperature: Shipping/Storage:
Relative Humidity: Shipping/Storgage:
Atmospheric Pressure: Shipping/Storage:
-40 to +70 ºC
10 to 100% RH
500 to 1060 hPa