User Manual H301-OLYMPUS-IX-SUSP REV 02 H301-OLYMPUS-IX-SUSP Compatible with the following XY stages Compatible with the following Okolab Controllers • • Olympus: IX-SUSP, GX-SVR, GX-SFR, BX3-SSU, H301-T-BL-PLUS IX2-SFR, IX2-SVL-2 • UNO-COMBINED-CONTROLLER • H401-T-DUAL www.oko-lab.com...
User Manual H301-OLYMPUS-IX-SUSP REV 02 Index Components and dimensions....................3 Sample Holders ........................3 2.1 Available Sample Holders ....................... 3 Available Lids ........................4 3.1 Sliding lid .......................... 5 3.2 Koehler Lid ........................5 3.3 Injection Lid ........................6 3.4 Laser Interlock Lid ....................... 7 Insertion of the Sample Feedback Temperature Sensor ..............
Sample Holder-Order Separately Integrated Humidty Module Chamber top view Chamber Base Figure 1. H301-OLYMPUS-IX-SUSP - Components and Dimensions. 2. Sample Holders Available Sample Holders The following sample holders are available. NOTE: Please contact info@oko-lab.com...
3. Available Lids The following heated glass Lids are available for H301-OLYMPUS-IX-SUSP: • H301-SLIDING-LID: For easy sample loading and pipetting – Included in code H301-OLYMPUS-IX-SUSP • H301-KOEHLER-LID-2*: Reduces chamber height to 23 mm and allows imaging under Koehler illumination •...
A in image 2. Open Sliding Lid to access screws labeled A in image 3. Sliding Lid Distance between focal plane and upper side of the glass lid Chamber top view Figure 3. H301-OLYMPUS-IX-SUSP (Sliding Lid comes as standard) Sliding Lid Screws M2x6 Figure 4. Assembly of the sliding lid Koehler Lid The Koehler Lid is a heated glass lid reducing the chamber’s height to 23 mm and allowing imaging under Koehler il-...
Distance between focal plane and upper side of the glass lid Chamber top view Figure 5.H301-OLYMPUS-IX-SUSP + H301-KOEHLER-LID-2 Injection Lid The Injection Lid is a heated glass lid with two small openings sealed with flexible plastic allowing injection or perma- nent access to the sample.
Distance between focal plane and upper side of the glass lid Chamber top view Figure 6. H301-OLYMPUS-IX-SUSP + H301-INJECTION-LID Laser Interlock Lid The Laser Interlock Lid is a heated glass lid with a safety switch connecting to the laser controller. It automatically turns the laser off when the lid is lifted.
H301-OLYMPUS-IX-SUSP REV 02 4. Insertion of the Sample Feedback Temperature Sensor Insert the Sample Feedback Temperature Sensor through the dedicated opening located in the H301-OLYMPUS-IX-SUSP (see Figure 8, Frontal and 3D views). Temperature Sensor dedicated opening Temperature Sensor dedicated opening 1.Frontal view...
User Manual H301-OLYMPUS-IX-SUSP REV 02 6. Working with 35 or 60 mm Petri Dish – Spacing Rings and Magnetic Locks Magnetic locks prevent movement of 35 and 60 mm dishes inside the sample holder. Figure 10 illustrates the available magnetic locks for 35 and 60 mm dishes. Threaded magnetic posts allow adjusting holder’s height.
User Manual H301-OLYMPUS-IX-SUSP REV 02 Ibidi μ-Dish35mm-low Grainer-Petri35x10 BD Falcon-35x10 Willco-35x10 Corning-35x10mm MatTek-P35G-X-14-X Figure 12. Ring selection scheme for 35 mm dish. Ibidi μ-dish 50mm-low MatTek-P50G-X-30-X BD Falcon 60x15 Grainer - Petri 60x15 Corning-60x15mm Willco60x15 Figure 13. Ring selection scheme for 60 mm dish.
Page 11
User Manual H301-OLYMPUS-IX-SUSP REV 02 NOTE: Magnetic locks are included with sample holder. Figure 14. Magnetic lock for 1'' x 3'' chamber slide. Figure 15. Magnetic lock for 1'' x 2'' chamber slide. To lock the slide, push simultaneously the buttons indicated with (3).
A single silicon tubing carries output gas from the Okolab Gas Controller to the H301-OLYMPUS-IX-SUSP. Silicon tubing connects to a gas input - brass opening - located on a side of the H301-OLYMPUS-IX-SUSP. See Figure 16. Connect by gently pushing silicon tubing onto brass opening.
User Manual H301-OLYMPUS-IX-SUSP REV 02 10. Connection of the Chamber with XY stage Follow the steps shown in the images Figure and listed below in order to correctly connect the chamber with XY stage. Place the chamber on the stage and tighten 4 captive screws (See Figure Image 1 and Image 2). Captive screws...
User Manual H301-OLYMPUS-IX-SUSP REV 02 11. Cleaning • Turn the system off and pull the mains plug out the socket • Remove the lid from the chamber main body, and keep it separate from the chamber main body while the chamber cools down.