User Manual H301-MINI Rev 02 H301-MINI Compatible with the following XY stages Compatible with the following Okolab Controllers • • Physik Instrumente: P-737.1SL/2SL/5SL H301-T-BL-PLUS • • Ludl 99A600 UNO-T • • It fits in any MW plate holder UNO-T-H-PREMIXED •...
User Manual H301-MINI Rev 02 Index Components and dimensions....................3 Sample Holders ........................4 Available Sample Holders ....................4 Insertion of the Sample Feedback Temperature Sensor ..............4 Insertion of Sample Holder into Chamber ................... 5 Working with 35 or 60 mm Petri Dish – Spacing Rings and Magnetic Locks ........... 5 Working with 1x3’’...
Heated glass lid with Indium Tin Oxide (ITO) conductive coating and embedded temperature sensor Distance between focal plane and upper side of the glass Sample Holder - Order Separately Chamber top view Chamber Base Figure 1. H301-MINI - Components and Dimensions. www.oko-lab.com...
Figure 2.Available sample holders. 3. Insertion of the Sample Feedback Temperature Sensor Insert the Sample Feedback Temperature Sensor through the dedicated opening located in the H301-MINI (see Figure 3, Frontal and 3D views). Temperature Sensor dedicated opening Temperature Sensor dedicated opening 1.Frontal view...
User Manual H301-MINI Rev 02 4. Insertion of Sample Holder into Chamber Sample holders fit into the chamber base and are held in place by magnets embedded within both chamber and hold- er. To introduce a sample holder with the proper orientation, match the red dot on the holder to the one on the chamber base, as illustrated in Figure 4.
Page 6
User Manual H301-MINI Rev 02 Figure 6. 1) Insertion of the ring, 2) insertion of the 35 mm dish, 3) insertion of the magnetic lock. NOTE: Spacing rings are included with sample holder. Ibidi μ-Dish35mm-low Grainer-Petri35x10 BD Falcon-35x10 Willco-35x10 Corning-35x10mm MatTek-P35G-X-14-X Figure 7.
User Manual H301-MINI Rev 02 6. Working with 1x3’’ and 1x2’’ chamber slides - magnetic locks Magnetic locks prevent movement of 1’’x 3’’ and 1’’x 2’’ chamber slides inside of the sample holder. NOTE: Magnetic locks are included with sample holder.
A single silicon tubing carries output gas from the Okolab Gas Controller to the H301-MINI. Silicon tubing connects to a gas input - brass opening - located on a corner of the H301-MINI. See Figure 11. Connect by gently pushing silicon tub- ing onto brass opening.
9. Connection of the Chamber with XY stage H301-MINI fits on the XY Stage without any screws. To fix the chamber on the stage, follow the steps shown in Figure Place the chamber on the stage (See Image 1 of Figure 13).
User Manual H301-MINI Rev 02 10. Cleaning • Turn the system off and pull the mains plug out the socket • Remove the lid from the chamber main body, and keep it separate from the chamber main body while the chamber cools down.