Operating Instructions; Blanket Control Features - Pedigo WARMWATCH P-2148 Operation & Care Manual

Combination blanket/fluid warming cabinet
Table of Contents

Advertisement

Operating Instructions

THERMOSTAT CONTROL AND L.E.D. DISPLAY
TIME
STATUS
.
.
.
INDICATOR L
E
D
180
.
.
.
L
E
D
DISPLAY
SET
/
ON
OFF
TEMPERATURE
The following refers to features that are available when the control
is powered on.
CONTROL PANEL BUTTONS
ON/OFF BUTTON
Press the ON/OFF button to power on the control. Press and
hold the ON/OFF button for 2 seconds to power the control
off. The status indicator L.E.D will illuminate according to the
power state.
UP ARROW / DOWN ARROW BUTTONS
Used to increase or decrease the temperature set-point.
Additionally used to set the current time, auto-start, and auto-
stop times. Continual pressure to a button will increase the
increments in which the values will change.
TIMER BUTTON
Blanket warmers have a timer button which is used to program
the time of day, as well as the automated start and stop times.
See "Setting the Time" instructions on this page.
L.E.D. DIGITAL DISPLAY
The control has a four-digit L.E.D. display. When the control
is activated, the display will show current temperature set-
point. When programming the timer, the display will show
hour and minutes.
L.E.D. DISPLAY STATUS INDICATORS
TIME
Illuminates while current time of day is displayed when
programming the TIMER mode.
START
Illuminates while the start time is displayed when programming
the TIMER Auto-On mode.
STOP
Illuminates while the stop time is displayed when programming
the TIMER Auto-Off mode.
LOCK
Illuminates when the lock feature is engaged.
POWER FAIL DETECTION
If the power were to fail for any reason while control is
powered on, the warmer will retain in memory its current
operating state. When the power is restored, the control will
alarm once and resume operating in its previously set mode,
but will alert the operator that such an event has occurred:
The ON/OFF status indicator will ash.
MM-1068 (MN-29617) • Rev 0 (10/15) • WarmWatch® Combination Blanket/Fluid Warmer • 11

Blanket Control Features

B L A N K E T
START
STOP
UP
STATUS
.
.
.
ARROW
INDICATOR L
E
D
.
F
-
POINT
AUTO
DOWN
LOCK
TIMER
ARROW
C O N T R O L
F E A T U R E S
Press the ON/OFF button once to acknowledge that the
power has been restored. The ON/OFF status indicator
will stop ashing. When pushing the ON/OFF button, the
display will indicate the time period of the outage in hours
and minutes (HH:MM), then return to the normal display and
previously set mode. Inspection of the product in the cavity
may be necessary.
TEMPERATURE FORMAT SELECTION
While the controller is in the off mode, press and hold the UP
ARROW button for 5 seconds. Press again to switch between
Fahrenheit (°F) or Celsius (°C).
CAVITY TEMPERATURE DISPLAY
To reference the cavity air temperature, push and hold
the TIMER and UP ARROW buttons. While holding both
buttons, the value in the display refers to the temperature
at the cavity sensor.
OPERATIONAL SOUND SELECTION
While the controller is in the off mode, press and hold the
DOWN ARROW button for 5 seconds. Press again to turn
the sound ON (I) and OFF (O).
CONTROL LOCK
The warmer control can be locked so that no changes can
be made to the temperature set-point or the mode selection.
Press the ON/OFF button and the UP arrow button at the
same time. The LOCK indicator will illuminate. Attempts to
operate the ON/OFF button, or to change the temperature
set-point will be unsuccessful. To unlock the control, press the
ON/OFF button and the DOWN arrow button at the same time.
The control will unlock and the LOCK indicator will extinguish.
SETTING THE TIME
NOTICE: If you do not wish to use the automated timer feature,
Press the TIMER button to illuminate the TIME indicator. The
current time of day will show on the display in 24-hour format
(HH:MM). Use the UP or DOWN arrow to set the display to
the current time. Press the TIMER button again to transition
to the START Auto-On Time. This is the time that the control
will automatically turn on when enabled. Adjust to desired start
time using the UP or DOWN arrow buttons. Press the TIMER
button again to transition to the STOP Auto-Off Time. This is
the time the control will automatically turn off when enabled.
Adjust to desired stop time. Press the TIMER button again
to enable the Auto-Timer feature. The TIME status indicator
L.E.D. will illuminate when the Auto-Time function is activated.
To disable, press and hold the TIMER button until the indicator
L.E.D. goes off.
NOTICE: When the Auto-On and the Auto-Off times are
then you do not need to set the time and this section
can be skipped.
equal, the unit will recognize the Auto-Off time
ONLY and the control will never turn on without
user intervention.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Warmwatch p-2140Warmwatch p-2145

Table of Contents