SYSTEM PROGRAM 2
SYSTEM PROGRAM 2 TIC + TOC e
RANGE CHANGE 1-2
RANGE CHANGE 1-3
RANGE CHANGE 2-1
SAMPLE IN
TIC ACID
BASE
BASE OXIDATION
TOC ACID
 RANGE CHANGE 1-2 defines the concentration levels for the
50.0mgC/l
BioTector to change its range automatically from range 1 to
range 2.
 The RANGE CHANGE 1-2 is typically the top concentration
point of TOC RANGE 1 in System Range Data menu by default.
 RANGE CHANGE 1-3 specifies the concentration level for the
150.0mgC/l
BioTector to change its range automatically from range 1 to
range 3.
 The RANGE CHANGE 1-3 is typically 50-75% of the top
concentration point of TOC RANGE 2 in System Range Data
menu by default.
 RANGE CHANGE 2-1 states the concentration level for the
40.0mgC/l
BioTector to change its range automatically from range 2 to
range 1.
 The RANGE CHANGE 2-1 is typically 20% less of the top
concentration point of TOC RANGE 1 in System Range Data
menu by default.
 SAMPLE IN identifies the amount of sample (in pulses) to be
2p
P-V
injected into the reactor at range 2.
 TIC ACID defines the amount of acid reagent (in pulses) to be
40p
injected into the reactor at range 2.
 BASE specifies the amount of base (in pulses) to be injected into
320p
the reactor for the BASE OXIDATION phase at range 2.
 BASE OXIDATION defines the oxidation time (135s by default),
135s
which can be programmed independently for range 2.
 If the Base Oxidation time is different than the one programmed
in System Program 1 menu, depending on the difference
between the two settings, the reaction time at range 2 will be
either longer or shorter than the reaction time displayed in
System Program 1 menu.
 TOC ACID states the amount of acid (in pulses) to be injected
280p
into the reactor for the TOC SPARGE phase at range 2.
Page 111