OUTPUT 1
DEFAULT STATE
STOP
FAULT
WARNING
NOTE
SAMPLER FILL
SAMPLER EMPTY
SAMPLER ERROR
SYNC
REMOTE STANDBY
MAN MODE TRIG
MAINT SIGNAL
TEMP. SWITCH
CAL
ZERO CAL
CAL SIGNAL
STREAM 1-3
MANUAL 1-3
SAMPLE STATUS 1-3
ALARM 1-3
CO2 ALARM 1-3
4-20mA CHNG
4-20mA CHNG 1-3
4-20mA READ
SAMPLE FAULT 1-3
COMPRESSOR
EQUALIZATION
SAMPLE PUMP REVERSE
N/D
DEFAULT STATE defines the idle state of the relay. N/D stands for "Normally De-
energized" relay by default and N/E stands for "Normally Energized" relay.
Output set to operate when the BioTector is stopped condition. Note that remote stand-
by is not considered as a stop condition.
Output set to operate on fault condition.
Output set to operate on warning condition.
Output activated when a notification is logged in the fault archive.
Fill signal, sent to the sampler, which remains on from start of sampler fill time to the
completion of the sample injection.
Empty signal, sent to the sampler, which is a pulse of 5 seconds duration and is
triggered after the Sample Pump reverse operation is complete.
Output when SAMPLER ERROR input signal is activated due to a sample error in the
BioTector Sampler.
Synchronization relay, use to synchronize the system with external control units.
Output when Remote Standby input is activated.
Indicates that manual reactions are activated and going to be executed regardless of
the activation is carried out manually on the keyboard or remotely from a system input.
*
Output when MAINTENANCE SWITCH input is activated.
Temperature Switch output activated when the temperature of the system increases
above the predefined system temperature control level (System Fan Control), which is
programmed as 20°C by default.
Calibration Valve used during span calibration/check reactions.
Calibration Valve used during zero calibration/check reactions.
*
Output set to operate on Zero/Span Check and Zero/Span Calibration reactions.
Output set for Stream Valves 1-3.
Output set for Manual Valves 1-3.
Digital output activated when BioTector Sample Sensor detects no sample
or when the sample quality is less than the default 75% threshold value for a specific
stream (i.e. when there is significant quantity of air bubbles in the stream/manual grab
sample lines).
Alarm relay, activated upon programmed alarm conditions for a specific stream.
CO2 Alarm relay, activated upon programmed CO2 alarm conditions for a specific
stream.
4-20mA output change flag relay, which is always activated for a period of 10s, when a
new result causes update of any of the 4-20mA output channels.
4-20mA output change flag relay, which is always activated for a period of 10s, when a
new result causes update of a given 4-20mA output channel for a specific stream.
This signal is used to indicate valid/stable values on 4-20mA output channels in 4-
20mA Stream & Full Multiplex operation modes.
Output when the external stream specific SAMPLE FAULT 1-3 input signal is activated.
Output when the compressor (Valve 1-J7 on Oxygen Controller Board) is activated.
Output set to operate from the beginning of the Sample Pump forward operation until
the Sample Valve activation.
Digital output activated during Sample Pump reverse operation.