Hach BioTector B3500e User Manual page 50

Online toc analyzer
Table of Contents

Advertisement

Figure 8 below illustrates the BioTector door clearance dimensions.
Figure 8 BioTector Door Clearance Dimensions
It is recommended that a clear space of at least 300mm is allowed around the analyzer from the right
and the left sides and from the top and the bottom. The recommended clear space in the front of the
analyzer is 1500mm for easy access to the system.
When BioTector is being mounted on a wall or a stand, the support has to be strong enough to carry
typically four times of the weight of BioTector (~180 kg).
The BioTector should be lifted applying a safe method in accordance with local regulations.
The minimum size of the bolts used to hold the BioTector in place should be M8.
Page 50

Advertisement

Table of Contents
loading

Table of Contents