Hach BioTector B3500e User Manual page 35

Online toc analyzer
Table of Contents

Advertisement

CONSUMABLES
Acid & Base:
Instrument Air:
Service:
ANALYSIS PARAMETERS
Oxidation Method:
TOC Measurement:
Measurement Terms:
Measured Components:
*
COD & BOD by correlation algorithm incorporating TOC measured results
Cycle Time:
MONITORING RANGES:
Standard Range
Accuracy and Repeatability
Detection Limit
Second Range
(Optional for Exceedance Tracking)
Range Selection:
SAMPLE & ENVIRONMENTAL CONDITIONS
Sample Volume:
Sample Inlet Pressure:
Drain Pressure:
Sample Inlet Temperature:
Sample Flow Rate:
Sample Particle Size:
Ambient Temperature:
Humidity:
Ingress Protection:
System Sound:
The manufacturer has a continuous research and development program. Specifications may
therefore be changed without notice. For specification updates, please contact the
manufacturer.
Typical Replacement Frequency & Consumption
4-8 weeks/19 Liters (based on typical ranges and 7.5 minute analysis cycle)
1.5 bar, - 20°C dew point (free of water, oil and dust)
Average consumption is less than 5.4 m
Filter pack is recommended and available to meet the air quality specification.
BioTector Air Compressor is available for air supply.
6 Monthly Intervals
Patented Two-Stage Advanced Oxidation Process using Hydroxyl Radicals
NDIR measurement of CO
TOC (Total Organic Carbon) as Non-Purgeable Organic Carbon (NPOC)
BioTector TIC&TOC mode measures NPOC.
TOC (NPOC)
TIC
TC
COD*
BOD*
from 6.5 minutes, depending on range and application
TOC
0-250mgC/l
±3% of reading or ±0.45mgC/l, whichever is greater
0.9mg/l
0-1000mgC/l
(
±4% of reading or ±2mgC/l, whichever is greater
Automatic or Manual (up to 2 ranges configurable)
Up to 3.0ml
Ambient (for applications with high sample pressure, sampling systems are
available)
Ambient (for applications with high drain pressure, optional systems are
available)
2°C – 60°C
(36°F - 140°F)
Minimum 100ml per sample
Up to 100 microns
5°C – 45°C
(41°F - 113°F)
Cooling and heating options are available.
5% - 85%, non-condensing
IP44, standard fan cooled, maximum ambient temperature 45°C (113°F)
IP54, air cooled, maximum ambient temperature 35°C (95°F)
IP54, vortex cooled, maximum ambient temperature 50°C (122°F)
< 60 dBa
/hour.
3
after oxidation
2
)
Page 35