regulator, which is dedicated for the vortex cooler only.
Connect the EXHAUST port with ¼" PFA tube to a safe and well ventilated area or to open
atmosphere. The tube must have no restrictions and it must be placed so that any condensation
and liquid buildup in the tubing is prevented. The maximum length of ¼" PFA tubing installed in
Exhaust line is 10 meters. If tubing longer than 10 meters is required, the use of a larger ID tubing
_____
or pipe is recommended.
The end of the exhaust tubing should have a slight downward slope and should be located in an
environment so that any condensation or liquid at the outlet of the tubing cannot freeze during
cold weather. See figure 16 in Section
5.4.2
Drain and Exhaust Connections
for details.
Page 68