5 Operation
3.
Press the
midity, dew point temperature, and mixing ratio.
•
Relative humidity: The
displayed in percent (%).
•
Dew point temperature: The
is displayed in ℃ or ℉, depending on the unit setting.
•
Mixing ratio: The reading is displayed in grains per pound (GPP) or grams
per kilogram (g/kg), depending on the unit setting.
5.4 IR temperature measurements
IR temperature measurements can be performed in all operating modes.
The meter is equipped with a laser pointer diode, which is used as a targeting
pointer for the IR temperature measurements. The target of the measurement
should be larger than the size of the laser beam spot. As the distance from an ob-
ject increases, the spot size of the area measured by the meter becomes larger.
The meter's field of view ratio is 8:1, meaning that if the meter is 8 cm (3.2") from
the target, the diameter (spot) of the object under test must be at least 1 cm
(0.4"). Refer to Figure 5.1.
Figure 5.1 IR spot-to-distance ratio
IR measurement notes:
•
The object under test should be larger than the than the size of the laser
beam spot.
•
If the surface of the object under test is covered with frost, oil, grime, etc.,
clean the surface before measuring.
•
If the surface of the object is highly reflective, apply masking tape or flat black
paint to the surface before measuring.
•
The meter may not make accurate measurements through transparent surfa-
ces such as glass.
#T559822; r.9065/9080; en-US
button repeatedly to cycle through the display of relative hu-
indicator is displayed and the reading is
indicator is displayed and the reading
14