Operating in Internal mode, the SR124 oscillator is programmed by
an internally-generated DC voltage source. Within the oscillator
frequency range (set by the Range knob), the oscillator frequency is
controlled with the front-panel control knob or remote command.
When configured for Rear VCO operation, the SR124 oscillator pro-
gramming voltage is directly controlled by the user through a rear-
panel BNC input. Input voltage from 0 to 10 V will set the oscillator
frequency between the lower and upper limits set by the range, with
an approximately linear transfer function.
The SR124 oscillator can lock to an external reference signal applied
to the Ext. In BNC connector. This input operates in two distinct
modes, sine input and TTL input (indicated by the TTL indicator
just above the connector).