4.4 Commands
OFST(?) g
Example:

4.4.9 Setup commands

KCLK(?) z
ALRM(?) z
SSET(?) z
RSET(?) z
SR124 Analog Lock-In Amplifier
Output o set magnitude
Set (query) the output o set magnitude to g , in percent full scale.
The default value is OFST 0.00.
The allowed range for g is 1000 g
OFST 300.253
Key clicks
Set (query) audible key clicks to z (OFF 0, ON 1) .
Note there is no corresponding front-panel method to access this
command; it is exclusive to the remote interface.
Audible alarms
Set (query) audible alarms to z (OFF 0, ON 1) .
Note that all sounds that are not "key clicks" are considered "alarms"
for the purpose of the ALRM command. There is no corresponding
front-panel methdo to access this command; it is exclusive to the
remote interface.
Save user settings
Save (query) the user settings to non-volatile block z (USER0 0,
USER1 1, . . . , USER8 8) .
The set version of SSET saves user settings to parameter block z.
The query version SSET? returns the most block name or number z
most recently saved into.
Recall user settings
Retrieve (query) the user settings to non-volatile block z (USER0 0,
USER1 1, . . . , USER8 8, DEFAULT 9) .
The set version of RSET retreives user settings from non-volatile
block z and reconfigures the SR124 to those settings. The query
form simply returns the name number of the most-recently retreived
memory block.
Note that RSET DEFAULT is equivalent to *RST.
1000 ( 10 full-scale).