Stanford Research Systems SR124 Operation And Service Manual page 75

Analog lock-in amplifier
Table of Contents

Advertisement

4.4 Commands
*STB? [i]
Example:
*SRE(?) [i,] j
Example:
*ESR? [i]
Example:
*ESE(?) [i,] j
Example:
*CLS
Example:
LEXE?
SR124 Analog Lock-In Amplifier
Status byte
Reads the Status Byte register [bit i].
*STB?
0
Service request enable
Set (query) the Service Request Enable register [bit i] to j .
*SRE 0,1
Standard event status
Reads the Standard Event Status Register [bit i].
Upon executing *ESR?, the returned bit(s) of the ESR register are
cleared.
*ESR?
64
Standard event status enable
Set (query) the Standard Event Status Enable Register [bit i] to j .
*ESE 6,1
ESE?
64
Clear status
*CLS immediately clears the ESR register, and the UNLOCK and
OVERLOAD bits in the Status Byte.
*CLS
Last execution error
Query the last execution error code. A query of LEXE? always clears
the error code, so a subsequent LEXE? will return 0. Valid codes are:
Value
Definition
0
No execution error since last LEXE?
1
Illegal value
2
Wrong token
3
Invalid bit
4
Queue full
5
Not compatible
4 – 21

Advertisement

Table of Contents
loading

Table of Contents