ATC
1
With the 1 point calibration, the acidic slope = the alkaline slope = -59.16 mV/pH is assumed.
With the 2 point calibration, the acidic slope = the alkaline slope.
With the 3 point calibration, independent acidic and alkaline values are determined.
Press the
9.2

Real-time clock ("CLOC")

The real-time clock is required for the chronological assignment of the calibra-
tion. You should therefore check the settings as required, particularly after
replacing the battery.
All manuals and user guides at all-guides.com
rH
mV
mV
pH
H
The date of the last calibration is shown in the main display.
The time of the last calibration is shown in the subsidiary dis-
play.
ok
CAL
AL Logg
pH
key to stop reading the calibration data.