KROHNE OPTISONIC 6300 F Quick Start page 3

Ultrasonic clamp-on flowmeter for liquids

Advertisement

3 Installation menu X
Measuring mode
Select menu

Press > 2.5 s
X1 language
X Installation
X2 GDC IR interface
X3 units
X4 pipe configuration
install transd. 1

1 shows when configuring a 2 pipe installation
11/2021 - 4008317002 - OPTISONIC 6300 F/W R02 en
Select Submenu
Functions

X3.1 size
X3.2 volume flow
X3.3 free unit
X3.4 [m3/s]*factor
X3.5 velocity
X3.6 density
X3.7 temperature
X4.1 number of pipes
select
X4.2 number of paths
X5.2 diameter
X5 pipe data
X5.3 pipe material
X5.4 VoS pipe material
X5.5 wall thickness
X5.6 liner material
X5.7 VoS liner material
X5.8 liner thickness
X5.9 fluid
X5.10 VoS fluid
X5.11 density
X5.12 glycol % vol.
X5.13 dynamic viscosity
X5.14 pipe temperature
X6 if pipe=2 at X4.1
X6._: identical to X5._
X7.1 transducer set
X7.2.1 calibration number
X7.2.2 number of traverses
X7
X7.2.3 mount transducers at
X7.2.4 act. flow, prelim.
X7.2.5 check signal
X7.2.6 actual distance
X7.2.7 optimize distance
X7.2.8 act.flow, prelim.
X7.2.9 path ready?
X7.2.11 end installation
Ready? or next transducer? /
install transd. 2 /
X8._ is identical to X7._
X9.1 Ta serial no.
X9.2 Ta calibration no.
X9
X9.3 Tb serial no.
transducer sets
X9.4 Tb calibration no.
X9.5 Tc serial no.
X9.6 Tc calibration no.


4 Quick Setup

Measuring mode
Select menu
Press > 2.5 s
A
Quick Setup


1
5
Status message
Tag number
Measured variable
Signal strength
Operating keys
Select Submenu


A1 Language
A2 Tag
A3.1 reset errors
A3 Reset
A3.2 totalizer 1
A3.3 totalizer 2
A3.4 totalizer 3
A4.1 measurement
A4 Analog Outputs
A4.2 unit
A4.3 range
A4.4 low fow cutoff
A4.5 time constant
A5.1 measurement
A5 Digital Outputs
A5.2 pulse value unit
A5.1 value p. pulse
A5.1 low fow cutoff
A6 GDC IR interface


Functions


6