Appendix C: Guidance and Manufacturer's Declaration
IMMUNITY
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions, and voltage
variations on power supply
input lines
IEC 61000-4-11
Note: U
is the A.C. mains voltage prior to application of the test level.
T
C-2
IEC 60601 TEST LEVEL A.C. Mains power voltage should be the typical
quality of a:
NA for power supply
lines.
±0.25 kV to ±1 kV
for input/output
lines.
COMPLIANCE LEVEL
NA for power supply
lines.
±0.25 kV to ±1 kV
for input/output
lines.
IEC 60601 TEST LEVEL
NA
COMPLIANCE LEVEL
NA
IEC 60601 TEST LEVEL
NA
COMPLIANCE LEVEL
NA
Pulse Oximeter for Sleep Screening Operation Manual
ELECTROMAGNETIC ENVIRONMENT
GUIDANCE
Commercial environment.