CADD-Legacy
Model 6400
Training Guide
WARNING: This pump delivers medication at a continuous rate in milliliters per 24 hours (ml/24hrs).
Programming the pump at a delivery rate other than what is prescribed will cause over- or under-delivery of medication.
6
Upstream Sensor (status only)
Verify setting is correct. Screen
shows whether sensor is turned On
or Off. To change setting, see BIOMED
FUNCTIONS.
5
Air detector (status only)
Verify setting is correct. Screen shows
whether sensor is turned On (High or
Low) or turned Off. To change setting, see
BIOMED FUNCTIONS.
1 Pump
®
1
Install new batteries in battery
compartment. Slide battery door closed.
In LLO
STOPPED
SCROLLING
Use the up and
down arrow keys
to increase or
decrease the value
on the screen.
4
Given (ml)
Displays total amount of medication
delivered since last cleared.
Press ENTER/CLEAR to clear value.
Rate is in
ml/24hrs
2
Reservoir Volume (ml)
Scroll to the volume contained
in the Medication Cassette
Reservoir. Press ENTER/CLEAR to
enter value.
3
Continuous Rate (ml/24hrs)
Select the infusion rate. Press
ENTER/CLEAR to enter value.