the power supply is 20.000V.
2)If sending command [:]VOLT:PROT? MAX, If the return command is 76.000, it means the maximum setup
voltage value of the power supply is 76.000V.

6.2.5 List Operation Related Command

Command [:]LIST:AREA <num>
1. This command is used to set the list location division mode.
Parameter: 1|2|4|8
1. Set 1 file of 200 list steps in the list operation
2. Set 2 files of 100 list steps in the list operation
4. Set 4 files of 50 list steps in the list operation
8. Set 8 files of 25 list steps in the list operation
Example:
If sending command sent [:]LIST:AREA 8, it means to set 8 files of 25 list steps.
2. Related command [:]LIST:AREA?, it means to query the location division mode of the list operation.
Return parameters: <num>
Command [:]LIST:RCL <num>
This command is used to recall the saved files from a certain location so as to make them executed in the list.
Parameter: 1~8
Return command: none
Example:
If sendin command [:]LIST:RCL 2, it means to recall the list file from the storage register location 2 and make
the file executed.
Command [:]LIST:COUNt <count>
1.This command is used to set the step number of list.
Parameter: count range : 1~200
Return command: none
Example:
If sending command [:]LIST:COUN 20, it means the step number of the list is set as 20.
2.Related command [:]LIST:COUNt?, it means to query the step number of the current list.
Return parameter:<count>
Command [:]LIST:MODE <mode>
1. This command is used to set the working mode of the list.
Parameters: mode has 3 types:
CONTinuous: continuous mode