Example:
If you send command [:]SYST: ERR?,then return command ought to be"0, 'No Error'", or "50, 'Error Para
Count'", or "70, 'Invalid Command'" etc..
Command [:]SYSTem:REM
Command [:]SYSTem:REM is used to set remote control mode, namely the PC control mode.
Return command: None
Example:
If you send command [:]SYSTem:REM
After sending command successfully, the power supply has been in the remote control mode , "REM"
characters appears in the right corner front panel and the front panel operation is not valid. And please return
the front panel control model by pressing the key Shift + the number 7, or sending the command [:] SYST:
LOC.
Command [:]SYSTem:LOC
Command [:] SYSTem:LOC is used to set the front panel control mode.
Return command: None
Example:
Sending command [:] SYST:LOC if the power supply in the remote control (PC) mode. After sending
successfully, the power has been in the front panel control mode while "REM"disappears from the right corner
of the front panel.
Command [:]SYSTem:SENSe <bool>
Command [:] SYSTem: SENSe <bool> is used to set the power is under remote measurement.
Return command: None
Parameters: bool-type parameter (0-OFF, 1-ON)
Example:
Sending command [:]SYST:SENS 1, if sending successfully, the power supply is set in the remote
measurement.

6.2.3 Measurement Command

Commands described in this section are under the root path [:]MEASure commands. The following commands
are used short-type mnemonic format.
Command [:]MEASure: VOLTage?
Command [:]MEASure: VOLTage? is used to read output voltage value.
Return command: Numeric
Return parameters Unit: V
Example:
Sending command [:]MEAS: VOLT?, if return command is 12.560, it means the present output voltage is