Thermo Scientific

Operation

Caution Exercise care when using acidic or caustic solutions as they will attack the
galvanized steel bath body if spilled into the bath. If spills do occur, the bath
liquid should be immediately drained and the unit thoroughly flushed. Spills and
condensation should be cleaned/removed from all metal surfaces after each use.
the user is reponsible for decontamination if hazardous material is spilled on or in
the bath.
When filling the bath with water, allowance must be made for the displacement
of water upon immersion of samples. Also make allowance for splashing of water
from tray oscillation movement.
Caution Water level requirements:
Minimum level - one inch of water should be maintained at all times. this level
will keep heaters covered and will help prevent heater failure.
Maximum level - water level should be kept lower than the tray shaft fitting.
water filled beyond this level could possibly seep into control panel and cause
damage.
To conserve energy, reduce evaporation and increase temperature control
accuracy, use the gable cover provided. Do not use aluminum foil as a cover, as it
may cause corrosion due to an electrochemical reaction.
1. Add distilled water to bath. Use water levels as instructed above.
2. Power up: Depress the power switch located at the lower right hand corner of
the bath. The unit is fully operational upon application of power. The display
will show the actual bath temperature.