List Of Diagnostic Events - Endress+Hauser HART Levelflex FMP56 Operating Instructions Manual

Guided level-radar level measurement in bulk solids
Hide thumbs Also See for HART Levelflex FMP56:
Table of Contents

Advertisement

Levelflex FMP56, FMP57
Endress+Hauser
à The message for the remedial measures for the diagnostics event opens.
2.
Press S + O simultaneously.
à The message about the remedial measures closes.
11.2.3

List of diagnostic events

Sensor element failures
Diagnostic event
Code
Description
F003
Broken probe detected
F046
Build-up detected
F083
Memory content
F104
HF cable
F105
HF cable
F106
Sensor
Electronic failures
Diagnostic event
Code
Description
F242
Software incompatible
F252
Modules incompatible
F261
Electronic modules
F262
Module connection
F270
Main electronic failure
M270
F271
Main electronic failure
F272
Main electronic failure
M272
F273
Main electronic failure
F275
I/O-Modul failure
F276
I/O-Modul failure
F282
Electronic memory
Maintenance instructions
1. Check map.
2. Check sensor.
Clean sensor.
1. Restart device.
2. Restore S-Dat data.
3. Change sensor.
1. Dry HF cable connection and check sealing.
2. Change HF cable.
1. Tighten HF cable connection.
2. Change HF cable.
1. Check probe isolation.
2. Change sensor.
Maintenance instructions
1. Check software.
2. Flash or change main electronics module.
1. Check electronic modules.
2. Change I/O or main electronic module.
1. Restart device.
2. Check electronic modules.
3. Change IO module or main electronics.
1. Check module connection.
2. Change electronic modules.
Change main electronic module.
1. Restart device.
2. Change main electronic module.
1. Restart device.
2. Contact service.
1. Emergency operation via display.
2. Change main electronics.
Change I/O module.
1. Restart device
2. Change I/O module
1. Restart device.
2. Contact service.
Trouble shooting
Error behavior
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Error behavior
Alarm
Alarm
Alarm
Alarm
Alarm
Warning
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
71