When mounting the electronics housing into a recess (e.g. in a concrete ceiling), observe a
minimum distance of 100 mm (4 inch) between the cover of the terminal compartment /
electronics compartment and the wall. Otherwise the connection compartment / electronics
compartment is not accessible after installation.
6.2.2
Applications with restricted mounting space
Mounting with remote sensor
The device version with a remote sensor is suited for applications with restricted mounting space.
In this case the electronics housing is mounted at a separate position from which it is easier accessible.
r
= 100 mm (4")
min
6 Nm
A
Angled plug at the probe
B
Angled plug at the electronics housing
• Levelflex version (see product structure):
Feature 600 "Probe Design", Option MB "Sensor remote, 3m/9ft cable, detachable+mounting
bracket" (® ä 19)
• A connecting cable is supplied with this device version
– Length: 3 m (9 ft)
– Minimum bending radius: 100 mm (4 inch)
• A mounting bracket for the electronics housing is supplied with this device version. Mounting
options:
– Wall mounting
– Pipe mounting; diameter: 42 to 60 mm (1-1/4 to 2 inch)
• The connection cable has got one straight and one angled plug (90°). Depending on the local
conditions the angled plug can be connected at the probe or at the electronics housing.
A
6 Nm
r
= 100 mm (4")
min
Mounting
B
6 Nm
6 Nm
A0014794
25