WARNING! Shorten the cheek and forehead supports of the tonometer only slightly at a
time to prevent the tonometer from getting too close to your eye.
WARNING! Do not connect anything to the tonometer's USB port but the USB cable
supplied with the tonometer.
WARNING! Keep the USB cable out of the reach of children and pets due to the risk of
strangulation.
WARNING! The tonometer's batteries are not rechargeable. Do not try to charge the
tonometer with USB chargers connected to a mains voltage.
WARNING! Do not connect the USB cable to the tonometer's USB port except when
uploading patient measurement data. Do not take any measurements when the USB
cable is connected.
WARNING! The tonometer should only be opened by qualified iCare service personnel.
The tonometer does not contain any user-serviceable parts, apart from the batteries and
the probe base. The tonometer does not require any routine servicing or calibration
other than changing the batteries at least annually and the probe base every six months.
If there is a reason to believe that the servicing of the tonometer is necessary, contact
the manufacturer or local distributor.
WARNING! The tonometer must not be repaired or re-assembled by any other than the
manufacturer or its authorized service center. If the tonometer is broken, do not use it.
Take it to an authorized iCare service center for repair.
WARNING! To avoid possible damage, keep the tonometer out of the reach of children
and pets. The probe base, battery cover, screws, collar, and probes are small objects and
may be accidentally swallowed.
WARNING! Do not change the batteries or the probe base when the USB cable is
connected.
WARNING! Servicing or maintenance actions must not be performed while the
tonometer is in use.
WARNING! The tonometer must be switched off when the probe base is changed.
WARNING! The probe base must be changed, not cleaned.
WARNING! Never immerse the tonometer in liquid. Do not spray, pour, or spill liquid on
the tonometer, its accessories, connectors, switches, or openings in the cover.
Immediately remove any liquid from the surface of the tonometer.
WARNING! Do not modify the tonometer in any way. Changes or modifications not
expressly approved by the manufacturer could void the user's authority to operate the
tonometer.
WARNING! Use of this equipment adjacent to or stacked with other equipment should
be avoided because it could result in improper operation. If such use is necessary, this
equipment and the other equipment should be observed to verify that they are
operating normally.
WARNING! Use of accessories, transducers, and cables other than those specified or
provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this equipment
and result in improper operation.
WARNING! Interference may occur in the vicinity of equipment marked with the
non-ionizing radiation symbol.
WARNING! Sources of power frequency magnetic field should be used no closer than 15
cm (6 inches) to any part of the tonometer, including the cables specified by the
manufacturer, to avoid the degradation of performance.
WARNING! Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30 cm (12 inches)
to any part of the tonometer, including the cables specified by the manufacturer, to
avoid the degradation of performance.
PRECAUTION! Read this manual carefully, as it contains important information about
using and servicing the tonometer.
PRECAUTION! Use the tonometer only for measuring intraocular pressure. Any other use
is improper. The manufacturer is not liable for any damage arising from improper use, or
for the consequences of such use.
PRECAUTION! Do not use the tonometer near inflammable substances, including
inflammable anesthetic agents.
PRECAUTION! Report any serious incidents related to the tonometer to your competent
health authority and the manufacturer or the manufacturer's representative.