Product description
548B
Screwdriver Torx TX8
17

Technical information

17.1

Technical description

WARNING! Do not modify the tonometer in any way. Changes or modifications not
expressly approved by the manufacturer could void the user's authority to operate the
tonometer.
NOTE! A separate service manual is available for service personnel.
Type: TA023
Dimensions (WxHxL): 50 mm x 94 mm x 152 mm (supports extended to
maximum) / 2.0" x 3.7" x 6.0" (supports extended to maximum)
Weight: 205 g (without batteries) / 0.45 lbs (without batteries), 300 g (with
4 x AA batteries) / 0.66 lbs (with 4 x AA batteries
Power supply: 4 x 1.5 V, AA, non-rechargeable batteries, alkaline LR6
Measurement range: 7 – 50 mmHg
Accuracy: ±1.2 mmHg (≤ 20 mmHg) and ±2.2 mmHg (> 20 mmHg)
Repeatability (coefficient of variation): <8%
Precision of display: 1 mmHg
Display unit: millimeters of mercury (mmHg)
Operation environment:
Temperature: +10 °C to +35 °C (50 °F to 95 °F)
Relative humidity: 30 % to 90 %
Atmospheric pressure: 800 hPa to 1,060 hPa
Storage environment:
Temperature: -10 °C to +55 °C (14 °F to 131 °F)
Relative humidity: 10 % to 95 %
Atmospheric pressure: 700 hPa to 1,060 hPa
Transport environment:
Temperature: -40 °C to +70 °C (-40 °F to 158 °F)
Relative humidity: 10 % to 95 %
Atmospheric pressure: 500 hPa-1,060 hPa
NOTE! It is recommended to let the tonometer temperature stabilize for about an hour to room
temperature before taking the tonometer to use after transport or storage.
The tonometer and its materials are compliant with RoHS Directive 2011/65/
EU. The tonometer and its parts are not made of natural rubber latex.
Method of sterilization of the probes: gamma-irradiation
Mode of operation: continuous
The serial number is located on the back of the tonometer. The LOT number
of the probes is on the side of the probe box and the blister packing. There
are no electrical connections from the tonometer to the patient. All the parts
of the tonometer are applied parts and the tonometer has BF-type electric
shock protection.
17.2

System requirements for iCare CLINIC

• Internet connection
• Minimum web browser versions: IE 11, Chrome (v 58), Firefox (v 53) and
Safari (5.1.7)
Weight
Dimensions
(in pounds)
WxHxL (in inches)
15 g
16 mm x 90 mm
(0.03 lbs)
(0.6" x 3.5")