Guidance and Manufacturer's Declaration IEC 60601-1-2:2014; Edition 4.0
iCare HOME2 tonometer (TA023) is intended for use in a home healthcare environment with
electromagnetic characteristics specified below. The user of the iCare HOME2 tonometer (TA023)
should assure that it is used in such an environment.
Immunity test
Conducted
disturbances
induced by RF
fields
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Electromagnetic Immunity
IEC 60601 test level
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur
radio
bands between 0,15
MHz and 80 MHz
80 % AM at
1 kHz
10 V/m
80 MHz – 2,7 GHz
Compliance level
3 V
6 V
10 V/m
Electromagnetic
environment-Guidance
WARNING!
Portable RF
communications
equipment (including
peripherals such as
antenna cables and
external antennas)
should be used no
closer than 30 cm (12
inches) to any part
of the iCare HOME2
(TA023), including
cables specified by the
manufacturer, to avoid
the degradation of
performance.
WARNING!
Portable RF
communications
equipment (including
peripherals such as
antenna cables and
external antennas)
should be used no
closer than 30 cm (12
inches) to any part
of the iCare HOME2
(TA023) including
cables specified by the
manufacturer, to avoid
the degradation of
performance.
Interference may
occur in the vicinity
of equipment marked
with the following
symbol: