Required Characteristics Of The It Network; Performance Data - Icare HOME2 Instruction Manual

Hide thumbs Also See for HOME2:
Table of Contents

Advertisement

from the tonometer. The measurement data can still be found from the
tonometer memory and transferred once the connection is re-established.
Failure or misconfiguration of the IT network may result in data not being
transferred.
17.6

Required characteristics of the IT network

The responsible organization is strongly recommended to maintain virus
protection up to date on the computers and mobile devices used. The
responsible organization is also recommended to install security updates to
the used web browsers, computers, and mobile devices when available.
17.7

Performance data

17.7.1
Clinical performance data
A clinical study was conducted to analyze the variability of the intraocular
pressure (IOP) self-measurements with the iCare HOME2 tonometer in
comparison to the variability of the IOP measurements with the reference
tonometer (iCare IC200, ANSI Z80.10-2014 compliant) over a wide range of
IOP measurement values.
The performance data was obtained from a clinical study. The study was
performed at East West Eye Institute, CA 90013, USA, and included 47
patients. All patients were found to be eligible for analysis. All the patients
were either diagnosed glaucoma patients or 'glaucoma-suspects'. A random
eye was selected as the study eye for each patient.
Safety: No adverse events (including corneal abrasions) were recorded in
the study population.
Results: The mean paired difference and standard deviation (iCare HOME2 -
iCare IC200) were 0.55 mmHg and 2.69 mmHg.
The iCare HOME2's variability (difference of repeat measurements) for each
patient was ~7.9% for all the IOP ranges.
Summary of study results (sitting and supine positions)
Group
N
≤ 16 mmHg
>16 to <23
mmHg
≥ 23 mmHg
Overall
PRECAUTION! Please be advised that the measurement results may vary between an
IOP self-measurement and an IOP measurement performed by a healthcare
professional.
In a clinical trial, the mean difference between the measurements performed by a
healthcare professional and an IOP self-measurement was -1.45 mmHg for sitting
position and 0.71 mmHg for supine position. The overall mean difference between self-
measured IOP values and healthcare professional measured IOP values was 0.55 mmHg.
HOME2
Reference,
IC200
Mean (SD)
Mean (SD)
24
15.78
(2.86)
13
20.17
(2.28)
10
23.88
(2.34)
47
18.72
(4.17)
Difference
Mean (SD)
14.86
-0.93
(2.9)
(2.75)
19.56
-0.6
(2.75)
(2.66)
24.33
0.44
(2.42)
(2.36)
18.17
0.55
(4.67)
(2.69)
95% CI
95% LOA
for Mean
for Mean
Difference
Difference
-1.38, -0.48
-6.32, 4.46
-1.2, -0.00
-5.81, 4.61
-0.17, 1.05
-4.19, 5.07
-0.86, -0.23
-5.82, 4.72
39

Advertisement

Table of Contents
loading

Table of Contents