troubleshooting and Errors

Fill or Drain Problems

4-2 Fill or Drain
Problems

4-2-1 Station cleaning

94
User Guide STP420D/STP420ES
A number of conditions can cause fill and drain problems:
There is insufficient reagent: Verify that the reagent bottles and the
paraffin tanks both have sufficient reagent.
Inadequate reagent from the external tank: When a message about
inadequate reagent coming from the external tank into the rotation
chamber appears, the reagent is already being pumped back into the
external tank.
There is insufficient pressure or vacuum: The STP420 cannot
create sufficient vacuum (for a fill) or pressure (for a drain) in the
chamber:
Verify that all reagent bottles are fully inserted in their correct
locations.
Ensure that the caps of the reagent bottles are tight.
Check if the lid of the chamber is properly closed.
There is a blockage in the air or reagent lines. Appropriate error
messages indicate the specific station that must be cleaned to
remove the blockage, for example:
"Cleaning Station X: Station X filling time (emptying time) is too long,
a station cleaning is recommended".
Station cleaning takes approximately 20 minutes:
1. Touch OK to confirm the error message.
2. Select Setup from the screen menu.
3. Select Cleaning from the submenu.
4. Select Station cleaning.
5. Select Start. A dialog box appears with the prompt: "Position X
contaminated. Please remove bottle X, check for paraffin residue and
clean if necessary".
6. Touch OK to acknowledge the prompt. A dialog box with a
command prompt: "Please insert the Xylolreinger bottle in position X
and put the bottle from the position X into a vacant position."
7. Follow the instructions on the screen and then click OK.