Environmental/Location Requirements - Konica Minolta DRYPRO SIGMA 2 Operation Manual

Table of Contents

Advertisement

Specifications

Environmental/Location Requirements

Temperature
Operating: 15 to 33 °C (59 to 91 °F)
Storage: -40 to 60 °C (-40 to 140 °F)
Relative Humidity
Operating: 20–80 % RH, non-condensing
Storage: 10–93 % RH, non-condensing
Altitude
-30 m (-100 ft) to 3,000 m (9,800 ft) above sea level
Surface Levelness
The surface where the imager is placed must be level within 1 °.
Environmental Effects
Acoustical noise:
Less than 48 dB at 1 m during idle or standby
Less than 75 dB momentary at 1 m during normal operation
Power Requirements
A power cord is provided with this equipment. All countries must use an
Agency-approved power cord with a plug type suitable for the country of use. Contact a
qualified dealer for help.
Connect the equipment to a power source that is suitable for the voltage and current
ratings shown on the rating label. The single-phase power source, with grounding, must
be provided within 2.5 m (8.0 ft) of the imager.
The wire must be insulation-rated for 600 V (ac). A dedicated line is recommended.