Basic Operating Tasks
The density values for the last test print are recorded in the Web Portal, and you can
manually record them on the film for charting purposes.
See the Mammography Quality Control Manual for more information.
QC Step Wedge Test Print
1. Press and hold the Calibrate symbol for the film cartridge (upper or lower, where the
mammography film is loaded) for three seconds.
2. The film prints with the four density steps. The display panel shows the Step 1 values.
3. Press the Calibrate symbol again to view the results of Step 2. Repeat for Steps 3–4.
Figure 12: In this example, the imager has printed a QC step wedge test print. The Step number
identifier displays at the upper left (1–4), and the density value displays on the right side (1.40 in
this example). Record the values, if desired, on the film.
4. Press the Pause symbol to return to normal operation.
AD3435_en | 2017-11-17