7.0

OPERATION

CAUTION:
EXERCISE CARE WHEN USING ACIDIC OR
CAUSTIC SOLUTIONS AS THEY WILL ATTACK
THE GALVANIZED STEEL BATH BODY IF SPILLED
INTO THE BATH. IF SPILLS DO OCCUR, THE
BATH LIQUID SHOULD BE IMMEDIATELY
DRAINED AND THE UNIT THOROUGHLY
FLUSHED. SPILLS AND CONDENSATION SHOULD
BE CLEANED/REMOVED FROM ALL METAL
SURFACES AFTER EACH USE.
7.01 When filling the bath with water, allowance
must be made for the displacement of water
upon immersion of samples. The maximum
liquid level is 1 inch below the top of the pan.
7.02 To conserve energy, reduce evaporation,
increase temperature control accuracy, and
reach/maintain 100°C, use the gable cover
provided. Do not use aluminum foil as a
cover, as it may cause corrosion due to an
electrochemical reaction.
1.
Add distilled water to bath. The water level
should be no closer than 1" from the top of the
bath when the bath is fully loaded. NOTE:
Use sterile distilled water only.
acceptable resistivity range is 50K to 1M
ohms (conductivity 1 to 20 mico Siemens).
The acceptable pH range is 7 to 9.
2.
Power up: Depress the power switch located
at the lower end of the control panel.
Immediately after turning on power, all of the
segments of the display will be on along with
the HEATER ON indicator. After 3 seconds,
the current set temperature will be displayed
and this will last for 5 seconds. Then the
display will change to the actual bath
temperature.
3.
To display the set temperature, quickly press
the /\ (up) or \/ (down) key. The current set
temperature will be displayed. After 5 seconds,
the display will return to the actual bath
temperature.
4.
To change the set temperature, continuously
press the /\ or \/ key until the desired set
temperature is displayed. When continuously
pressing either the /\ or \/ keys, the display
changes slowly at first but after 1 degree has
passed, the display begins to change rapidly.
After the desired set temperature is achieved,
then press the ENTER key. The display with
the new set temperature will flash three times
and then it will return to the bath temperature.
If the ENTER key is NOT pressed, the
expected new set temperature will not be
accepted and the water bath will control at the
previous set temperature.
5.
The water baths are calibrated at the factory
for use over a wide range of temperatures.
Due to non-linearities in the control system, it
may be necessary to make the display match
a calibrated thermometer's reading, even
though the difference might be only a few
tenth's of a degree.
The calibrate function should be used only to
match a STABLE bath's actual temperature
to the calibrated thermometer.
To perform calibration, press the CALIBRATE
key. The display changes to CAL, then flashes
The
back to bath temperature. This occurs four
times, if no other keys are pressed, then it
resumes displaying the bath temperature.
After inserting a calibrated thermometer into
the holder and letting it stabilize for a minimum
of 15 minutes, note its reading.
Press the CALIBRATE key once again and
while the display is flashing, press either the
/\ or \/ key to make the display match the noted
thermometer reading. After the desired display
is achieved, press the ENTER key. The display
flashes three times with the adjusted reading,
then stops flashing. The bath is now calibrated
more accurately for the control set point.
EXPLOSION, IMPLOSION OR THE RELEASE
OF TOXIC OR FLAMMABLE GASES ARISING
FROM THE MATERIAL BEING HEATED IS THE
SOLE RESPONSIBILITY OF THE USER.