CR Mammography | Introduction to the CR Mammography System | 27

Safety Directions

For general safety directions, refer to the Agfa HealthCare System Safety User
Manual, document 3100.
For product safety directions, refer to the User Manuals listed in 'System
Documentation'.
There are some good working practices that should be applied when operating
the CR Mammography System:
Related Links
System Documentation
Warning: If Micro Calcification Enhancement (MCE) is
applied, two instances of the image are available on the
PACS workstation: the original image and the MCE
enhanced image. Final diagnosis must be performed on the
original image.
Warning: MCE can enhance the noise of underexposed
images.
Caution: If the cassette is dropped before the exposure, open the
cassette and close it again to make sure for correct alignment to
the thorax side before the new exposure. For more information,
refer to the CR Plates and Cassettes User Manual (only
applicable for CR MM2.0 and CR MM3.0 cassettes and plates).
Caution: When using annotation boxes, be aware that the
diagnostic information can be possibly overlaid.
Caution: Lead markers for laterality (left or right) indication
shall be used. Lead markers should be arranged towards the
corners opposite of the chest wall and outside of the breast area.
Caution: It is important that the same image plate is always used
with the same cassette.