CR Mammography | Introduction to the CR Mammography System | 21
Caution: These above mentioned alignments are mandatory
because they were used for validation of the powder-based CR
Mammography Systems. In addition the EUREF based
alignments are in-line with the EUREF recommendations for
digital mammography.
For Mammography Systems based on CR HM5.0 (DX-M Digitizers)
Caution: For the EUREF based alignment, higher doses up to the
maximum AGD for DX-M may be used (target AGD for DX-M
with HM5.0 Image Quality Optimized) if a further improvement
of the image quality is desired.
Caution: If the CR HM5.0 is used besides the CR MM3.0R (or CR
MM2.0 and MM3.0) on the same modality, two separate AEC
channels of the X-ray device must be used and adjusted.
Caution: When using the CR HM5.0 instead of the CR MM3.0R
(and vice versa) on one X-ray device channel, a new AEC
calibration must be performed.