18 | CR Mammography | Introduction to the CR Mammography System
Print Layouts
For NX CR Workstation standard layouts are used.
These layouts optimize the hanging order of the films of the left and right
breast on the light box, by minimizing the borders at the thorax side of both
images.
CR Mammography Printer
Hardcopies with a maximum optical density of at least 3.6 are recommended
to display the image details properly.
Diagnostic Workstation
The softcopy reading station shall be installed and configured in presence of
the CR Mammography specialist. In case of deviation from these
recommended settings, Agfa does not guarantee optimal performance of the
system.
Softcopy reading stations of diagnostic quality require a dual head five mega-
pixel mammo display system.
2344Q EN 20180406 1001
Warning: For Diagnostic mammography use, true size
printing is essential. In this case, use only the proper
dedicated mammography printer layouts. If you use other
printer layouts, diagnostic information can be lost.