10 | CR Mammography | Introduction to the CR Mammography System
available formats are
18x24cm and
24x30cm
CR Workstation
The CR MM2.0 Mammo plates and cassettes, CR MM3.0 Mammo plate and
cassette, CR MM3.xR Mammo plate and cassette and CR HM5.x Mammo
detector are generically referred to as 'plates and cassettes'.
Combination restrictions:
Digitizer support for detectors, plates and cassettes:
• CR 25.0 and CR 75.0 support both CR MM2.0 Mammo Plates and
Cassettes and CR MM3.0 Mammo Plates and Cassettes.
• CR 35-X and CR 85-X supports CR MM3.0 Mammo Plates and Cassettes
only.
• CR 30-Xm supports CR MM3.0T Plates and Cassettes only.
• DX-M supports CR HM5.0 or CR MM3.0R Plates and Cassettes.
• Mixed use of different mammo cassettes and plates is not allowed. In
particular the mixed use of CR MM3.0R and CR HM5.0 on DX-M is not
allowed.
2344Q EN 20180406 1001
CR MM3.0
CR MM2.0 CR HM5.0
NX for Mammography
CR MM3.0R CR MM3.0T