7. Copy the desired files to your computer.
A list of patient sessions data can be found in the sessions folder.
A list of test logs can be found inside the TESTS folder. The files are
named with the Timestamp of the test start.
In the figure below, a data log from a 6_MWT is shown
Figure 29 - 6_MWT data log
In the top 4 rows is shown the parameters for test setup. The next
rows show the actual test data provided with one record per second
(rows). It only records during the test Duration time and not the
Warmup nor the Recovery time.
To create your own test:
Go to the root of the drive and find the file named TESTS.
Go to the next row (row 5) and provide it with a Name, Duration,
Warmup Cooldown (Recovery) and a valid Profile name (for instance
'Walking', 'COPD_NORM' or a user defined one). To change an
existing test, modify values for Duration, Warmup or Cooldown.
D 8.4.1.20 O
matic User Manual - IFU
64
2