Figure 11: Patient ID change
d. Set a desired Alarm Delay for losing signal on the pulse
oximeter sensor.
This is to accommodate differences in patient mobility. Some
patients are very immobile and in critical condition, where a low
alarm delay is necessary, while others can be moving around a
lot, thus more suited for a higher delay.
e. Select a profile.
Profiles are a set of parameters tied together to avoid repetitive
patient setups. Standard profiles can be set up in the
administrator menu and selected in the patient menu. However,
the active profile can also be customized if needed. The actual
active profile is stated at the top of the screen.
Note: Table A.4 in appendix A shows default factory set profiles.
d. Place the sensor on the patient's finger
e. Apply a nasal catheter or an oxygen mask on the patient.