Flow Error
Oxyg
en
Power
/Batte
ry
D 8.4.1.20 O
matic User Manual - IFU
2
O
2
Check
status
Battery
symbol
alarm
constant
Check
status
Battery
symbol
alarm
blinking
Power
symbol
alarm
blinking
Check
status
Incorrect O
flow
2
into device
Oxygen control
error
"Battery failed"
under device
status
Low battery
Critically low
battery
No power
Memory fail
Check oxygen
source. If
problem isn't
from source
device needs
repair
Check oxygen
source. If
problem isn't
from source
device needs
repair
Restart device,
if alarm persist
device needs
service
If device has
been without
external
power,
recharge the
battery. Else
have the
battery
changed
If device has
been without
external
power,
recharge the
battery. Else
have the
battery
changed. It is
not
recommended
to use device
in this alarm
state
Connect power
adapter to
power source
and to O
matic
2
Restart device,
if alarm persist
device needs
service
57