O2MATIC QUICK GUIDE
Get started quickly with O2matic
Applies to O2matic OMC PC-100 & PRO 100 with software version 1.5.0

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the OMC PC-100 and is the answer not in the manual?

Questions and answers

Summary of Contents for O2matic OMC PC-100

  • Page 1 O2MATIC QUICK GUIDE Get started quickly with O2matic Applies to O2matic OMC PC-100 & PRO 100 with software version 1.5.0...
  • Page 2 2. Apply the Nonin sensor on the finger of the patient. 3. Apply mask or nasal catheter. NOTE: Make sure the switch on the back of the device is turned on. O2matic PRO will turn on with the latest saved settings.
  • Page 3 Change patient or profile 1. Press ”Patient”. 2. Check if the profile settings are correct - See figure 2. 3. You may now: a) Press ”Select” followed by using the up/down buttons to select another profile. b) Change the response time of how fast the device alarms on loss of signal. (”Oximeter Signal Delay”).
  • Page 4 Flow: 0 - 15 l/min Based on BTS guidelines. WALKING Used for 6 minutes walking test and other SpO2: 90 - 94% mobility tests. In this case, O2matic should be attached to a rollator and run in battery mode. Flow: 0 - 15 l/min...