SETUP AND ADJUSTMENT

The EE210 is ready to use and does not require any configuration by the user. The factory setup of EE210 corresponds to the type
number ordered. For ordering guide please see data sheet at www.epluse.com/ee210. If needed, the user can change the factory
setup by using the USB configuration adapter (HA011066) and the EE-PCS, Product Configuration Software.
The user can change the scaling and the analogue output signal, the digital settings and perform RH and T adjustment.
Note: The EE210 may not be connected to any additional power supply when using the USB configuration adapter HA011066.
PC
EE-PCS PRODUCT CONFIGURATION SOFTWARE
1. Download the EE-PCS Product Configuration Software from www.epluse.com/configurator and install it on the PC.
2. Connect the E+E device to the PC using the USB configuration adapter.
3. Start the EE-PCS software.
4. Follow the instructions on the EE-PCS opening page for scanning the ports and identifying the connected device
5. Click on the desired setup or adjustment mode from the main EE-PCS menu on the left and follow the online
instructions of the EE-PCS

MAINTENANCE

Humidity calibration and adjustment:
Depending on the application and the requirements of certain industries, there might arise the need for periodical humidity calibration
(comparison with a reference) or adjustment (bringing the device in line with a reference).
Calibration and adjustment at E+E Elektronik
Calibration and/or adjustment can be performed in the E+E Elektronik calibration laboratory. For information on the E+E
capabilities in ISO or accredited calibration please see www.eplusecal.com.
Calibration and adjustment by the user
Depending on the level of accuracy required, the humidity reference can be:
Humor 20 Humidity Calibrator, please see www.epluse.com.
Omniport30 handheld device, please see www.epluse.com/omniport30.
Calibrated salt solutions, please see www.epluse.com/ee210.
Temperature calibration and adjustment:
Due to the outstanding protection of the Pt1000 temperature sensing element integrated in the E+E RH/T sensing element, a drift of
the T measurement is rather unlikely. If adjustment seems necessary, although the user can perform a one or two point T adjustment
with USB configuration adapter and EE-PCS against a reference of his choice, it is highly recommended to return the device to the
manufacturer for this. The reasons rest on the difficulty of an accurate T calibration in the air. The calibration shall take into account
the self-heating of EE210 with closed enclosure, in its real mounting position and in continuous operation, the impact of the output
current and of the probe orientation to the self-heating, as well as the cooling effect of the air circulation in a climate chamber possibly
used for calibration.
When employed in dusty, polluted environment:
The filter cap shall be replaced once in a while with an E+E original one. A polluted filter cap causes longer response time of the
device.
If needed, the sensing head can be cleaned. For cleaning instructions please see www.epluse.com/ee210.