Hinweise zur elektromagnetischen
Verträglichkeit (EMV)
Recommended separation distances between portable and mobile RF communications equipment and the
KARL STORZ flexible CMOS Video-Cysto-Urethoscopes 11272 V/VU in conjunction with Monitor 8402 ZX
The flexible CMOS Video-Cysto-Urethoscope 11272 V/VU in conjunction with Monitor 8402 ZX is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The user of the unit can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the unit as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum output
power of transmitter
150 kHz to 80 MHz
[W]
D
0.01
0.1
0,1
0.4
1
1.2
10
3.7
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.
Appendix
Electromagnetic Compatibility (EMC)
Information
Table 206
Separation distance d [m]
according to frequency of transmitter
80 MHz to 800 MHz
D
0.1
0.4
1.2
3.7
Vedlegg:
Merknader om elektromagnetisk
kompatibilitet (EMC)
800 MHz to 2.5 GHz
F
0.2
0.7
2.3
7.4