6 Cleaning the Transmitter Probe

CAUTION!
a well ventilated area. Observe all necessary safety precautions
when handling cleaning solvents.
To clean the moisture probe, you will need the following items:
Three glass (
2 containers, each with approximately 300 ml of reagent-grade
hexane or toluene
1 container, with approximately 300 ml of distilled (
deionized) water.
CAUTION!
submerge the probe. Do not place the transmitter module into
the solvents. You should only insert the sensor portion of the
transmitter into the solvents.
Rubber or latex gloves.
An oven set at 50°C ± 2°C (122°F ± 3.6°F), for drying the probe
A 1-1/8 in. wrench
Refer to the following sections to properly clean the probe.
VeriDri™ User's Manual
Be sure to perform the probe cleaning procedure in
metal) containers with the following solvents:
NOT
Make sure the containers are deep enough to
Operating Instructions
NOT
9