6 Cleaning the Transmitter Probe
CAUTION!
a well ventilated area. Observe all necessary safety precautions
when handling cleaning solvents.
To clean the moisture probe, you will need the following items:
•
Three glass (
•
2 containers, each with approximately 300 ml of reagent-grade
hexane or toluene
•
1 container, with approximately 300 ml of distilled (
deionized) water.
CAUTION!
submerge the probe. Do not place the transmitter module into
the solvents. You should only insert the sensor portion of the
transmitter into the solvents.
•
Rubber or latex gloves.
•
An oven set at 50°C ± 2°C (122°F ± 3.6°F), for drying the probe
•
A 1-1/8 in. wrench
Refer to the following sections to properly clean the probe.
VeriDri™ User's Manual
Be sure to perform the probe cleaning procedure in
metal) containers with the following solvents:
NOT
Make sure the containers are deep enough to
Operating Instructions
NOT
9