Installation Guidelines; Mounting The Probe - GE DewPro MMY31 Installation And Operation Manual

Trace moisture transmitter
Hide thumbs Also See for DewPro MMY31:
Table of Contents

Advertisement

DewPro MMY31
2.0

Installation Guidelines

2.1 Mounting the Probe

Preferably, mount the probe vertically such that the sensor tip points down. Mount
the probe in a G 1/2 thread connection or with a flange. Be certain that the tip of the
probe does not touch the inside wall of the pipe.
Adjust the ferrule (nylon or stainless steel*) for a probe insertion length of 1"
(minimum) and tighten the compression fitting as follows:
Hand-tighten the nut. Then, using a wrench, tighten the nut one and one-
half turns.
Now the fitting is tight and can withstand pressure to 17 bar (250 psi)
(provided a stainless steel ferrule is used)
GE General Eastern
GENERAL EASTERN
The Humidity Experts
MMR 31
MODEL:
SERIAL NO:
SUP PLY:
9...32 VDC
DEWPO INT RANG E:
P max:
P SIG
Minimum 25mm
Installation and Operation Manual
Fig. 3
6