6.2 Soaking the Sensor and Shield

CAUTION!
solvents. You should only insert the sensor portion of the
transmitter into the solvents. Also, do not allow the sensor to
come into contact with the surfaces of the cleaning containers or
any other hard surface.
While wearing protective gloves, complete the following steps:
Place the sensor in the first container of hexane or toluene and
1.
allow it to soak for 10 minutes.
Remove the sensor from the hexane or toluene and soak it in the
2.
container of distilled water for 10 minutes.
Remove the sensor from the distilled water and soak it in the second
3.
(clean) container of hexane or toluene for 10 minutes.
Remove the sensor from the hexane or toluene and set it aside until
4.
the shield has completed its cleaning cycle.
Clean the shield by repeating steps 1 to 3 above. To ensure the
5.
removal of any contaminants that may have become embedded in
the porous walls of the shield, swirl the shield in the solvents during
the soaking procedures.
Remove the shield from the hexane or toluene.
6.
Carefully replace the shield over the exposed sensor,
7.
touching the sensor
Place the sensor with the installed shield in an oven set at
8.
50°C ± 2°C (122°F ± 3.6°F) for 24 hours.
VeriDri™ User's Manual
Do not place the transmitter module into the
.
Operating Instructions
without
11