Operating Instructions
6.3 Evaluating the Probe
To evaluate the probe, complete the following steps:
Re-connect the cable to the transmitter module and measure the
1.
ambient dew point. Make sure you measure the same ambient air as
measured in step 2 on page 10.
Compare the two ambient air readings. If the new ambient air
2.
reading is within ±2°C (±3.6°F) of the first reading, the cleaned
probe is properly calibrated and may be reinstalled. If the two
readings not within that tolerance, proceed to step 3 below.
If the probe is still not reading the ambient air accurately, repeat the
3.
cleaning procedure using soaking times that are 5 times the
previous cleaning sequence, until two consecutive ambient air
readings are within ±2°C (±3.6°F).
If the above cleaning procedure does not result in accurate readings, contact
GE for assistance.