(1) Above dimensions subject to change without notice.
(2) Weight is without preamplifier assembly. Add 4 lbs. for each preamplifier assembly.
1.2
Specifications for "PT" Meters
"PT" Meter, rim and blade type, when installed with meter tube and calibrated on water.
straightening sections and
Linearity:
3" and larger:
•
•
•
•
•
Repeatability:
±0.02% at any point throughout the extended minimum to extended maximum flow range
Pressure Drop:
At maximum flow of normal flow range on water 4 psi
Pressure Range:
ANSI 150-1500# depending upon size
Viscosity:
High viscosities will reduce flow range of meter. Consult Daniel Measurement engineers.
Pickup Coil:
Nominal open circuit peak to peak voltage output:
Minimum flow
Maximum flow
Table 1-2:
Item
Meter body
Flanges (slip on)
Suspension
Installation and maintenance instructions
±0.25% normal linearity
±0.5% over extended minimum to extended maximum flow range
±0.15% premium linearity
±0.10% premium + linearity
±0.07% premium ++ linearity
Materials of construction
Blade
4.0v
0.3v
Standard
304SS
Carbon steel
304SS
Dimensions and specifications of the meter
Rimmed
5.0v
0.1v
Optional
316SS
304SS or 316SS
316SS
3