Summary of Contents for Emerson Daniel Ultrasonic 100 Series
Page 1
Installation procedure P/N 3-9000-749, Rev G January 2018 ™ Daniel Ultrasonic Series 100 Option Board Field...
Page 2
+44 (0)1786.433400 +44 (0)1786.433401 Middle East Africa (Dubai, UAE) +971 4 8118100 +971 4 8865465 Daniel Measurement and Control, Inc. (Headquarters) 11100 Brittmoore Park Drive Houston, TX 77041 USA http://www.emerson.com Email • Customer Service: Daniel.SystemSales@Emerson.com • Customer Support: Daniel.SystemSales@Emerson.com •...
Page 3
Signal words and symbols Pay special attention to the following signal words, safety alert symbols and statements: Safety alert symbol This is a safety alert symbol. It is used to alert you to potential physical injury hazards. Obey all safety messages that follow this symbol to avoid possible injury or death.
Page 4
Important safety instructions Daniel Measurement and Control, Inc. (Daniel) designs, manufactures and tests products to function within specific conditions. Because these products are sophisticated technical instruments, it is important that the owner and operation personnel must strictly adhere both to the information printed on the product and to all instructions provided in this manual prior to installation, operation, and maintenance.
Page 5
Product operation (Personnel): • To prevent personal injury, personnel must follow all instructions of this manual prior to and during operation of the product. • Follow all warnings, cautions, and notices marked on, and supplied with, this product. • System should be designed to avoid over pressure conditions or exceeding maximum safe flow rate if meter losses measurement.
Page 6
Notice THE CONTENTS OF THIS PUBLICATION ARE PRESENTED FOR INFORMATIONAL PURPOSES ONLY, AND WHILE EVERY EFFORT HAS BEEN MADE TO ENSURE THEIR ACCURACY, THEY ARE NOT TO BE CONSTRUED AS WARRANTIES OR GUARANTEES, EXPRESSED OR IMPLIED, REGARDING THE PRODUCTS OR SERVICES DESCRIBED HEREIN OR THEIR USE OR APPLICABILITY. ALL SALES ARE GOVERNED BY DANIEL'S TERMS AND CONDITIONS, WHICH ARE AVAILABLE UPON REQUEST.
Page 7
Warranty and Limitations 1. LIMITED WARRANTY: Subject to the limitations contained in Section 2 herein, Daniel Measurement & Control, Inc. (“Daniel”) warrants that the licensed firmware embodied in the Goods will execute the programming instructions provided by Daniel, and that the Goods manufactured by Daniel will be free from defects in materials or workmanship under normal use and care and Services will be performed by trained personnel using proper equipment and instrumentation for the particular Service provided.
Contents Contents Part I Plan Chapter 1 Introduction ........................3 General information of this document ..................3 Requirements to start ........................3 1.2.1 Required parts .......................3 1.2.2 Required tools .......................5 Part II Install Chapter 2 Installation of the Option Board ..................9 Instructions before opening the unit ....................9 Instructions to install the Option Board ..................9 Appendices and reference...
Page 10
Contents Daniel Ultrasonic Series 100 Option Board Field...
Introduction Introduction Topics covered in this chapter: • General information of this document • Requirements to start General information of this document Welcome to the Daniel Ultrasonic Series 100 Option Boards Field Installation Procedure manual. This manual has been designed to provide you with a step-by-step set of instructions for installing one of the Series 100 Option boards.
Page 14
Introduction Quantity Description 2-4-3400-098 2-position plug for Field Con- nection Board when using the Series 100 Option Board Kits 2-position plug for Field Con- nection Board when using the Series 100 Plus HART Option Board Kits 2-4-3400-099 3 position plug for Field Con- nection Board 2-4-3400-085 Wiring diagram cards for upper...
Installation of the Option Board Installation of the Option Board Topics covered in this chapter: • Instructions before opening the unit • Instructions to install the Option Board Instructions before opening the unit To prevent electrostatic damage to the electronic boards, always use a Ground Strap while handling the circuit boards.
Page 20
Installation of the Option Board Figure 2-1: Open electronics enclosure Daniel Mark III Gas Ultrasonic Meters Series 100 Option Board: Requires version 1.30 firmware or later. Series 100 Plus HART Option Board: Requires version 1.60 firmware or later. Daniel 3800 Liquid Ultrasonic Meters Series 100 Option Board: Requires version 1.00 firmware or later.
Page 21
Installation of the Option Board Push the locking mechanism on both card guides to the unlocked position as shown. Figure 2-2: Unlock CPU board Slide the CPU board out of the enclosure. Use the 1/8" flat-head or flat-blade screwdriver to unscrew the two Acquisition board cable screws. Disconnect the cable and remove the CPU board from the enclosure.
Page 22
Installation of the Option Board Figure 2-4: Prepare Option board Daniel Ultrasonic Series 100 Option Board Field...
Page 23
Installation of the Option Board Figure 2-5: Option Board Connectors/Switches/LEDs Installation procedure...
Page 24
Installation of the Option Board Figure 2-6: HART Option Board Connectors/Switches/LEDs Make any desired changes to the switch settings on the Option board before installation. Use the Wiring Diagram Card C-15470 provided as a reference. Install the Option board on the CPU board making sure that the connector on the bottom of the Option board mates to the connector on top of the CPU board.
Page 25
Installation of the Option Board Figure 2-8: Secure Option board If the Wiring Diagram Card C-15470 attached to the enclosure is older than the one provided in the Option Board Kit, replace it with the newer revision. a. Use a hex driver to remove the nut that attaches the Wiring Diagram Card. Remove the old card and install the new card provided with the Option Board Kit.
Page 26
Installation of the Option Board Figure 2-9: Install wiring Option card Reattach the Acquisition board cable to the CPU board and secure it by tightening the two retaining screws. Figure 2-10: Attach acquisition cable to CPU board Daniel Ultrasonic Series 100 Option Board Field...
Page 27
Installation of the Option Board Slide the CPU board back onto the card guides and seat it in the Field Connection board. Push the locking mechanism on both card guides to the Locked position as shown to secure the CPU board. If the mechanisms do not move over, make sure the CPU board is fully seated in the enclosure and try again.
Page 28
Installation of the Option Board Figure 2-12: Install plugs Reinstall the end caps on each end of the upper electronics enclosure. a. Reapply power to the unit. Figure 2-13: Install end caps Daniel Ultrasonic Series 100 Option Board Field...
Engineering Drawings Appendix A Engineering Drawings DE-21056 Mark III Ultrasonic System Wiring, Ultrasonic Flowmeter Model 3400, 3410 and 3420 (Sheets 1-3) Installation procedure...
Page 30
Engineering Drawings Daniel Ultrasonic Series 100 Option Board Field...
Page 35
Daniel Measurement and Control, Inc. ("Daniel") is an Emerson Automation Solutions business unit. The Daniel name and logo are trademarks of Daniel Industries, Inc. The Emerson logo is a trademark and service mark of Emerson Electric Co. All other trademarks are the property of their respective...