INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE
5.
Select a sidestream airway adapter. Make sure the windows
are clean and dry. Use the sidestream sampling tubing to con-
nect the airway adapter to the input connector on the etCO2
module or pod.
6.
Connect a nasal sampling cannula to the sidestream airway
adapter. If using dehumidification tubing, place in line.
7.
Align the marks on the bottom of the adapter and the bottom
of the CAPNOSTAT sensor and snap the adapter into the sen-
sor until you hear a click.
8.
Place cannula on patient appropriately and secure cabling out
of patients' way.
MAINSTREAM MONITORING SETUP
1.
Use Rotary knob to highlight etCO2 parameter box, click.
2.
Use Rotary knob to highlight Measurement Mode, click.
3.
Use Rotary knob to highlight Main, click.
Select a mainstream airway adapter. Make sure the windows
4.
are clean and dry. Replace the adapter if necessary.
Align the marks on bottom of the adapter with the sensor and
5.
snap firmly into place.
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
|31