INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE
CONFIGURE PARAMETER ALARM LIMITS
1.
Press Alarm Limits fixed key
2.
Use Rotary Knob to highlight the parameter you wish to con-
figure, i.e. NBP, click
3.
Use Rotary knob to adjust Alarms ON/OFF, click
4.
Use Rotary knob to adjust Upper limit, or Lower Limit, click
5.
Use Rotary knob to adjust Archive setting
Off – no event is stored and no recording is generated.
o
Store – stores the event for later review.
o
Record – generates a timed recording.
o
Str/Rec – generates a timed recording and stores the
o
event.
6.
Use Rotary knob to adjust Alarm Volume, OFF, 10-100 %.
Alarm Volume can only be set to OFF in OR mode or if monitor
is connected to the Central Station.
7.
Use Rotary knob to highlight ST menu icon, click to access ST
parameter setup.
CONFIGURE ARRHYTMIA ALARM LIMITS
1.
Press Alarm Limits fixed key.
2.
Use Rotary knob to highlight ARR menu icon at bottom right
corner of table, click.
3.
Use Rotary knob to highlight ARR parameter you wish to con-
figure, click.
4.
Use Rotary knob to adjust each function you wish to modify,
click.
RELEARNING THE ECG WAVEFORM
QRS complex is saved as a reference template. Relearn is suggested
every 8 hours, after repositioning leads, and when QRS diagnosis is
in question.
1.
Access Alarm Limits table, see above
2.
Use Rotary knob to highlight Arr Menu, click
3.
Use Rotary knob to highlight Relearn, click
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
|16
© 2014 Draeger Medical, Inc.