INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE
optimize breath count.
Manual
-
The monitor detects breath based on the size of the
o
waveform.
After turning on Manual mode adjust waveform size.
o
Intended for adult or pediatric patients whose breathing
o
patterns show excessive variation; or for neonates whose
breathing rhythms tend to be irregular, and whose signals
may not be reliably detected in Auto mode.
TO ADJUST RESPIRATORY WAVEFORM SIZE
Waveform size must be adjusted in Manual mode as it sets the
breath detection threshold based on the waveform size. Adjusting
the size in Auto mode is for size only.
1.
Use Rotary knob to highlight Respirations parameter box, click
2.
Use Rotary knob to highlight Size, click
3.
Start at 5%, gradually increase by 5% increments to optimize
count rate. (5-100%)
TO CHANGE RESPIRATIONS LEAD
1.
Use Rotary knob to highlight Respirations parameter box,
click.
2.
Use Rotary knob to highlight RESP Lead, click,
3.
Use Rotary knob to select lead, I or II, click.
Lead II placement of electrodes for the Adult and Neonate
TO RELEARN RESP WAVEFORM
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
|22
© 2014 Draeger Medical, Inc.